Fastgen Corporation

Cardiac-3 Rapid Test

INTENDED USE:   The Cardiac-3 Rapid Test is intended for the in vitro diagnostic use in the qualitative  detection  of cardiac Troponin I, Myoglogin, and CK-MB in human serum samples.

 

Features and Advantages of Cardiac-3 Rapid Test

  • Highly sensitive chromatographic immunoassay* detects the levels of all three critical cardiac markers for the diagnosis or exclusion of acute myocardial infarction (AMI)

  • Incorporates measurement of CK-MB used in the WHO criteria for MI diagnosis

  • Provides complete biochemical data in one test

  • Rapid test format delivers results in minutes

*US FDA 510(K) approved pending 

Performance Characteristics

  • Cut-off positivity concentrations -

Troponin I 1.5 ng/ml
Myoglobin 100 ng/ml
CK-MB 7.0 ng/ml
  • Accuracy

Troponin I

 

Negative
(0 ng/ml)

Troponin I
(0.3-0.8 ng/ml)

Troponin I 
(≥1.1 ng/ml)

# Samples 105 23 67
Negative 104 23 0
Position 10 0 67
Agreement 99.1% 100% 100%

Myoglobin

 

Negative
(0 ng/ml)

Myoglobin
(13-66 ng/ml)

Myoglobin 
(
>100 ng/ml)

# Samples 70 50 101
Negative 70 42 0
Position 0 8 101
Agreement 100% 84% 100%

CK-MB

 

Negative
(0 ng/ml)

CK-MB
(0.8-6.9 ng/ml)

CK-MB 
(
>7.5 ng/ml)

# Samples 79 30 58
Negative 79 26 0
Position 0 4 58
Agreement 100% 86.7% 100%